Purpose: Circulating water baths with high precision temperature uniformity and control for working with enzymes and handling of semen. The temperature control should be microprocessor controlled and uniformity of temperature of whole bath be controlled by precision circulation.Specifications for Circulating water bath
1) Working Temperature Ambient + 5 to 150 oC or more
2) Temperature stability 0.03 oC or better
3) Heating Capacity 2 KW
4) Pump capacity flow rate 15 L/min or better
5) Pump capacity pressure 0.35 bar or better
6) Bath chamber capacity 16 Liters or more
7) Filling Volume 12-15 liters
8) Early warning system for low liquid level
9) The pump flow rate can be manually control by the user for desire flow rate.
Additional requirements:
a) Warranty of the equipment: 2 years
b) IQ OQ PQ documentation for equipment certification should be provided
c) We provided Manufacturing, testing certificates.
D.D.R. INTERNATIONAL
All Rights Reserved.(Terms of Use) Developed and Managed by Infocom Network Private Limited. |